Providing a Total Solution for all your X-Ray System needs
| Gemini 0150F10 | |||
| Basic | |||
| Output Power | 150 W | ||
| Tube Voltage Range (kV) | 25-100 | ||
| Tube Current Range (mA) | 0.05-6 | ||
| Mechanical | |||
| Tank Dimension (mm) | 308 × 205 × 150 | ||
| Tank Weight (kg) | 17.2 | ||
| Inverter Dimension (mm) | 301 × 152 × 55 | ||
| Inverter Weight (kg) | 1.5 | ||
| Electrical | |||
| Tube Voltage Ripple | <0.5% (RMS) | ||
| Tube Voltage Accuracy | ± 1% | ||
| Tube Voltage Rising Time (ms) | 500 | ||
| Tube Current Accuracy | ± 1% | ||
| Power Supply | |||
| Mains Input |
1Φ, 90~264VAC / 50-60 Hz |
||
| Cooling Fan Input |
1Φ,220VAC or 110VAC/50-60Hz |
||
| X Ray Tube / Tank | |||
| Focal Spot Size (mm) | 0.5 | ||
| Beam Geometry | 80° × 20° Fan | ||
| Operational | |||
| Interface | RS232 | ||
| Software | GUI service tool | ||
| Operating Mode | Continuous Exposure Mode | ||
| Thermal | |||
| Max. Tank Heat Storage | 450 kHU / 338 kJ | ||
| Tank Heat Dissipation (W) | 150 | ||
| Environmental | |||
| Operating Temperature (℃) | 0-40 | ||
| Storage & Transportation Temperature (℃) | -20-70 | ||
| Altitude | 0 to 4000m | ||
| Operating Humidity(Non-condensing) | 5-95 | ||
| Storage & Transportation Humidity (Non-condensing) | 5-95 | ||
* This document cannot be considered to be a contractual specification.